2
Installation
CPI Canada Inc.
Page 2-16
Rev. H Indico 100 Series Service Manual Ch # 740895-04
2.12.0 PROGRAMMING
THE
LOW
SPEED
STARTER
This section applies only to units fitted with the low speed starter.
PLEASE BE SURE TO READ AND UNDERSTAND THIS SECTION FULLY BEFORE PROCEEDING.
The low speed starter boost and run voltages are fixed at nominally 240 VAC and 50 VAC respectively.
The phase shift capacitor is fixed at 25 uF. Therefore, the low speed starter is capable of operating the “R”
type stator tubes listed in table 2-1 only.
If the desired tube type is not listed, please contact CPI product support for assistance.
WARNING:
240 VAC IS PRESENT ON THE LOW SPEED STARTER BOARD AT ALL TIMES THAT THE
GENERATOR IS SWITCHED ON. TAKE APPROPRIATE PRECAUTIONS WHEN SERVICING
THIS BOARD
2.12.1 Low
Speed
Starter
Tube
Select
Table
TABLE 2-1: TUBE TYPES (LOW
SPEED STARTER)
TUBE TYPE
(HOUSING)
TUBE TYPE
(INSERT)
Comet DO7
25/50
Ω
stator
DX7
Comet DO9
25/50
Ω
stator
DX9 0.6/2.0
DX9 1.2/2.0
Comet DO9
25/50
Ω
stator
DX91HS
DX92HS
DX93HS
Comet DO10
25/50
Ω
stator
DX10HS 0.6/1.0
DX10HS 1.0/2.0
Comet DX10
25/50
Ω
stator
DX101HS
DX104HS
DX105HS
Comet
XSTAR8
XSTAR74
25/50
Ω
stator
XST-8
XST-74
Gilardoni
Rotagil S/AS
AR11-30
AR30-60
Picker
PX1300
3” anode
Std “R stator”
PX1302
PX1312
Picker
PX1400
4” anode
Std “R” stator
PX1429
PX1431
PX1436
PX1482
Philips
ROT350
ROT351
RO 17/50
SRO 22/50
SRO 33/100