3D
AEC Calibration
CPI Canada Inc.
Page 3D-8
Rev. D Millenia / Indico 100 Series Service Manual Ch # 740904-06
3D.2.0 PRECALIBRATION
SETUP
3D.2.1 AEC
Setup
Worksheet
Before continuing, it is suggested that a copy of the table below be filled in with all required information. Refer
to the example AEC setup worksheet on the next page.
FUNCTION
RECEPTOR 1
RECEPTOR 2
RECEPTOR 3
RECEPTOR 4
Film/Screen 1.
2.
3.
Nominal optical density:
Grid ratio/SID:
Min - max kVp range:
±
Density steps
∗
Density dose
change
+%:
(per step)
∗
-%:
Chamber type:
Regulatory AEC dose
requirements?
Is film processing maintained?
Assigned receptor name:
Are all cassettes similar?
Additional notes:
Additional notes:
Table 3D-1: AEC setup worksheet
ALL RECEPTORS MUST HAVE THE SAME NUMBER OF DENSITY STEPS AND THE SAME DENSITY DOSE
CHANGE PER STEP (DENSITY SETTINGS ARE COMMON TO ALL FILM SCREENS AND RECEPTORS).