CPI Canada Inc.
AEC Calibration
3D
Millenia / Indico 100 Series Service Manual Ch # 740904-06 Rev. D
Page 3D-9
3D.2.1 AEC Setup Worksheet (Cont)
Note: The example below is supplied for reference only. It does not represent an actual installation.
FUNCTION
RECEPTOR 1
RECEPTOR 2
RECEPTOR 3
RECEPTOR 4
Film/Screen 1.
Lanex/reg Lanex/reg Lanex/reg PMT/I.I.
2.
Lanex/med Lanex/chest
3.
Nominal optical density:
1.2 1.1 1.4
Grid ratio/SID:
12:1 8:1 10:1 10:1
Min - max kVp range:
60 - 120
65 - 140
80 - 110
70 - 120
±
Density steps
±
8
±
8
±
8
±
8
Density dose
change
+%: 12.5 12.5 12.5 12.5
(per step)
-%: 6.25 6.25 6.25 6.25
Chamber type:
Ion Solid
state
Ion PMT
Regulatory AEC dose
requirements?
Yes Yes Yes Yes
Is film processing maintained?
Yes Yes Yes Yes
Assigned receptor name:
Table Wall SFD Digital
Are all cassettes similar?
Yes Yes Yes N/A
Additional notes:
Additional notes:
Table 3D-2: Sample AEC setup worksheet