CPI Canada Inc.
ABS Calibration
3E
Millenia / Indico 100 Series Service Manual Ch # 740904-07 Rev. D
Page 3E-9
3E.2.5 DC Proportional
1. Dress the signal cable from the camera to allow it to connect to J7 on the generator interface board.
Wire to J7 as per figure 3E-3. Alternately, the DC proportional signal output may be connected to BNC
connector J8 on the generator interface board.
2. Ensure that there is sufficient slack in the cable such that the cabinet door does not strain this cable
when opening and closing. Secure the cable in place such as to prevent mechanical stress on the
connections.
3. Position the jumpers on the generator interface board as per table 3E-2 for Proportional DC,
selecting the correct configuration depending on whether the input is at J7 or J8 and whether the
polarity is negative or positive.
3E.2.6 Composite Video
1. Dress the composite video output from the camera to allow it to connect to J8 on the generator
interface board.
2. Ensure that there is sufficient slack in the cable such that the cabinet door does not strain this cable
when opening and closing. Secure the cable in place such as to prevent mechanical stress on the
terminations.
3. Position the jumpers on the generator interface board as per table 3E-2 for Composite Video,
selecting the correct configuration depending on whether the video is 75
Ω
terminated or high
impedance (it must be determined beforehand whether the 75
Ω
termination must be made at the
generator).