3B
System Interfacing
CPI Canada Inc.
Page 3B-6
Rev. E Indico 100 Series Service Manual Ch # 740895-05
3B.3.1 Inputs
•
All inputs are opto coupled, select inputs may be configured to use an ex/- 24 VDC source or
may be configured to accept a closed dry contact.
•
Note the following inputs and jumper configurations per the table below:
NOTE:
The Indico 100 generator is shipped from the factory with all inputs configured for dry
contact inputs.
JUMPER CONFIGURATION (INPUTS):
ROOM INTERFACE BOARD
GENERATOR INTERFACE BOARD
JUMPER CONFIGURATION
TB1 PINS 4 & 5 (SPARE)
JW7 PINS 1-2, 3-4 = DRY CONTACT INPUT
JW7 PINS 2-3 = 24 VDC EXTERNAL INPUT
TB2 PINS 6 & 7 (COLLIMATOR)
JW9 PINS 1-2, 3-4 = DRY CONTACT INPUT
JW9 PINS 2-3 = 24 VDC EXTERNAL INPUT
TB2 PINS 4 & 5 (BUCKY CONTACTS)
JW10.PINS 1-2, 3-4 = DRY CONTACT INPUT
JW10 PINS 2-3 = 24 VDC EXTERNAL INPUT
TB3 PINS 6 & 7 (TOMO EXPOSURE)
JW3 PINS 1-2, 3-4 = DRY CONTACT INPUT
JW3 PINS 2-3 = 24 VDC EXTERNAL INPUT
TB3 PINS 4 & 5 (REMOTE TOMO SELECT)
JW2 PINS 1-2, 3-4 = DRY CONTACT INPUT
JW2 PINS 2-3 = 24 VDC EXTERNAL INPUT
TB4 PINS 8 & 9 (THERMAL SWITCH 1)
DRY CONTACT INPUT ONLY
TB4 PINS 6 & 7 (THERMAL SWITCH 2)
DRY CONTACT INPUT ONLY
TB4 PINS 4 & 5 (ROOM DOOR INTLK)
DRY CONTACT INPUT ONLY
TB5 PINS 11 & 12 (MULT. SPOT EXPOSURE)
JW6 PINS 1-2, 3-4 = DRY CONTACT INPUT
JW6 PINS 2-3 = 24 VDC EXTERNAL INPUT
TB6 PINS 9 & 10 (REMOTE EXPOSURE)
JW15 PINS 1-2, 3-4 = DRY CONTACT INPUT
JW15 PINS 2-3 = 24 VDC EXTERNAL INPUT
TB6 PINS 7 & 8 (REMOTE PREP)
JW14 PINS 1-2, 3-4 = DRY CONTACT INPUT
JW14 PINS 2-3 = 24 VDC EXTERNAL INPUT
TB6 PINS 3 & 4 (I.I. SAFETY)
JW8 PINS 1-2, 3-4 = DRY CONTACT INPUT
JW8 PINS 2-3 = 24 VDC EXTERNAL INPUT
TB6 PINS 5 & 6 (REMOTE FLUORO EXP)
DRY CONTACT INPUT ONLY