3C
Programming the Generator
CPI Canada Inc.
Page 3C-16
Rev. B Millenia / Indico 100 Series Service Manual Ch # 740904-05
3C.5.3 Receptor Setup (cont)
5.
Certain selections are toggled (press the selection button again to change the
value), other parameters must be selected via the adjacent selection button.
The values are then changed using the + and - buttons.
6.
When finished setting the parameters and/or defaults for the current receptor,
press the << or EXIT button(s) as required to return to the GEN
CONFIGURATION menu.
7.
Reselect RECEPTOR SETUP, then select the next receptor to be programmed.
8.
When finished programming all receptors, return to the GEN
CONFIGURATION menu as per step 6.
3C.5.4 I/O Configuration
The I/O CONFIGURATION menus allow programming the states of exposure for the inputs and the
outputs on the room interface board.
* INPUTS [sym] *
REMOTE EXP : ...__...__... STANDBY STATE
REMOTE PREP: ...__...........
REMOTE FL. EXP: .............---
CONSOLE EXP: ...---...---...
EXIT
⇑
>>
* INPUTS [sym] *
CONSOLE PREP: ...---......... STANDBY STATE
TOMO EXP: ........ ---...
REM. TOMO SEL.: ---............
I/I SAFETY: ...---...------
<<
⇑
>>
* INPUTS [sym] *
COLL. ITLK: ...---...------ STANDBY STATE
BUCKY CONTACTS: .........---...
SPARE: ------...------
THERMAL SW 1: ------...------
<<
⇑
>>
* INPUTS [sym] *
THERMAL SW 2: -----...----- STANDBY STATE
DOOR ITLK: ----- ...-----
MULTI SPOT EXP: ---............
<<
⇑
>>