1C
Preparing for Installation
CPI Canada Inc.
Page 1C-2
Rev. D Indico 100 Series Service Manual Ch # 740895-02
1C.1.0 INTRODUCTION
The following items must be considered before installing your generator:
•
Power level of your generator.
•
Power line requirements.
•
Ground requirements.
•
Physical placement of the generator.
•
Environmental requirements for the generator.
•
Cable runs from the generator to all room components: tables, buckys, X-ray
tubes etc.
1C.2.0 GENERATOR POWER REQUIREMENTS
1C.2.1 30 kW Single Phase
Line Voltage
230 VAC
±
10%, 1
∅
.
Line Frequency
50/60 Hz.
Momentary Current
200 Amps.
Standby Current
5 Amps.
1C.2.2 37.5 kW Single Phase
Line Voltage
230 VAC
±
10%, 1
∅
.
Line Frequency
50/60 Hz.
Momentary Current
240 Amps.
Standby Current
5 Amps.
1C.2.3 30 kW Three Phase
Line Voltage
400 VAC
±
10%, 3
∅
(for 400 VAC input generators).
480 VAC
±
10%, 3
∅
with optional line adjusting transformer,
(for 400 VAC input generators).
480 VAC
±
10%, 3
∅
(for 480 VAC input generators).
*
SEE NOTE AT THE END OF 1C.2.7
Line Frequency
50/60 Hz.
Momentary Current
80 Amps/phase at 400 VAC.
67 Amps/phase at 480 VAC.
Standby Current
5 Amps.