5
Troubleshooting
CPI Canada Inc
Page 5-4 Rev. B Millenia / Indico 100 Series Service Manual Ch # 740904-09
5.2.3 Error
Messages
(Cont)
ERROR CODE
MESSAGE
PROBLEM
ACTION
E008 KV/MA
FAULT
Power
supply
has
detected
a fault in the kV or mA
output during an exposure
and immediately terminated
the exposure. This may be
caused by arcing in the X-
ray tube, arcing of the HV
cables, or HT tank.
1. If arcing of the X-
ray tube is
suspected, check
condition of tube.
The X-ray tube
may be damaged
or simply require
“seasoning”.
Refer to chapter 6
for tube
seasoning
procedure.
2. If failure of HT
tank is suspected,
contact product
support.
E009
PS NOT READY
Power supply is not ready to
make an exposure.
Retry exposure.
E011
HIGH MA FAULT
Generator CPU detected
mA greater than allowed
tolerance.
Recalibrate X-ray
tube.
E012 LOW
MA
FAULT Generator
CPU
detected
mA less than allowed
tolerance.
Recalibrate X-ray
tube.
E013
MANUAL TERMIN
Operator released exposure
switch during exposure.
1. Re-take exposure
if necessary.
2. Check for faulty
switch contacts or
wiring.
E014
AEC BUT ERROR
AEC exposure exceeded
allowed back up time.
1. Check exposure
technique
settings.
2. Check that correct
AEC chamber is
energized.
E015
AEC BU MAS ERR
AEC exposure exceeded
allowed back up mAs.
1. Check exposure
technique
settings.
2. Check that correct
AEC chamber is
energized.