CPI Canada Inc.
Installation
2
Indico 100 Series Service Manual Ch # 740895-04 Rev. H
Page 2-13
2.9.5 Generator
Mains
Connection
WARNING:
TO AVOID ELECTRICAL SHOCK, ENSURE THAT THE AC MAINS DISCONNECT IS LOCKED
IN THE OFF POSITION, AND THAT ALL MAINS CABLES ARE DE-ENERGIZED BEFORE
CONNECTING TO THE GENERATOR.
Refer to chapter 1C for generator power and generator power line requirements.
1.
Pass the AC mains cable through the access hole located at the lower rear of the generator.
2.
Use an appropriate cable clamp to secure the mains cable at the cabinet entrance.
3.
Temporarily remove the safety cover from the main fuses. Strip sufficient cable jacket to allow the
ground wire to reach the main ground connector located at the left side of the main fuse block.
Refer to figure 2-7.
4.
Connect the ground wire to the chassis ground connector, and connect the mains wires to the
terminals on the bottom of the main fuseholder (3 wires for 3 phase systems, 2 wires for single
phase systems). Be sure to replace the main fuse safety cover after all connections are made and
properly tightened.
5.
DO NOT SWITCH ON MAINS POWER AT THIS TIME.
Figure 2-7: Generator mains connections