3D
AEC Calibration
CPI Canada Inc.
Page 3D-36
Rev. D Millenia / Indico 100 Series Service Manual Ch # 740904-06
3D.8.0 RLF
COMPENSATION
(Cont)
The following screens are used for RLF compensation
* AEC CALIBRATION *
FILM SCREEN 1
DENSITY SETUP
FILM SCREEN 2
FILM SCREEN 3
EXIT
*TUBE? 0%HU AEC CAL, F/S 1*
50KV: 84.0:
85KV: 42.8
55KV: 78.0:
95KV: 38.0
65KV: 66.0:
+
75KV: 54.0
-
3200MS << >>
*TUBE? 0%HU AEC CAL, F/S 1*
110KV: 34.0:
130KV: 28.0:
RLF COMPENSATION
+
MULT. SPOT COMP: 0%
-
3200MS <<
*TUBE? 0%HU LF COMP, F/S 1*
50MSEC: 0%
500MSEC: 4%
1000MSEC: 0%
+
-
3200MS <<