5
Troubleshooting
CPI Canada Inc
Page 5-6 Rev. B Millenia / Indico 100 Series Service Manual Ch # 740904-09
5.2.3 Error
Messages
(Cont)
ERROR CODE
MESSAGE
PROBLEM
ACTION
E032 CONSOLE
COMM
ERR
Generator
has
detected
error in communication to
console.
1. Check console
cable for damage
and proper
connection.
2. Turn power off and
then on to reset
generator.
E033 GEN
BATTERY
LOW
Generator
detects
lithium
battery voltage is low.
Replace lithium
battery.
E034
+12VDC ERROR
+12VDC rail is out of
tolerance.
Check +12VDC rail.
E035
-12VDC ERROR
-12VDC rail is out of
tolerance.
Check -12VDC rail.
E036
+15VDC ERROR
+15VDC rail is out of
tolerance.
Check +15VDC rail.
E037
-15VDC ERROR
-15VDC rail is out of
tolerance.
Check -15VDC rail.
E038 CAL
DATA
ERROR
Generator
detects
corrupt
calibration data.
Re-calibrate X-ray
tube(s).
E039
AEC DATA ERROR
Generator detects corrupt
AEC data.
Reprogram AEC data
or set factory
defaults.
E040 FLUORO
DATA
ERROR
Generator
detects
corrupt
Fluoro data.
Reprogram fluoro
data or set factory
defaults.
E041
REC DATA ERROR
Generator detects corrupt
receptor data.
Reprogram receptor
data or set factory
defaults.
E042 TUBE
DATA
ERR Generator
detects
corrupt
tube data.
Reprogram tube data
or set factory
defaults.
E043
KV ERROR
KV detected in non x-ray
state.
Switch OFF
generator. Prevent
further use of
generator. Call
product support.
E044 COMM
ERROR Received
communication
message not valid and
ignored.
Reset error.
E045
NOT SUPPORTED
Received message valid,
but not supported by this
system.
Reset error.
E046
MODE INHIBITED
Received message valid,
but not allowed during
present state.
Reset error.
E047
FL TIMER LIMIT
Fluoro Timer has exceeded
time limit.
Reset Fluoro timer.