CPI Canada Inc.
Programming the Generator
3C
Millenia / Indico 100 Series Service Manual Ch # 740904-05 Rev. B
Page 3C-19
3C.5.5 AEC Setup
The AEC SETUP menu allows the setting of AEC parameters for each AEC channel.
* AEC SETUP*
CHANNEL: 1
CHAMBER TYPE: ION
LEFT FIELD: YES
FILM SCREEN 1 : YES
CENTER FIELD: YES
FILM SCREEN 2 : YES
RIGHT FIELD: YES
FILM SCREEN 3 : YES
EXIT
Definitions of AEC SETUP parameters as used in this section.
•
CHANNEL
Selects the AEC channel to be programmed.
•
LEFT FIELD
Enables or disables the left field for the selected AEC channel
(NO is disabled).
•
CENTER FIELD
As above but for center field.
•
RIGHT FIELD
As above but for right field.
•
CHAMBER TYPE
Selects ION CHAMBER or SOLID STATE (S/S) chamber for
the selected AEC channel.
•
FILM SCREEN 1
Enables or disables the selection of FILM SCREEN 1 for that
AEC channel (NO is disabled).
•
FILM SCREEN 2
As above but for film screen 2.
•
FILM SCREEN 3
As above but for film screen 2.
Use these steps to perform the AEC SETUP.
Step
Action
1.
From the GEN CONFIGURATION menu select AEC SETUP.
2.
Select the AEC channel to be setup. Pressing the CHANNEL button will scroll
through the available AEC channels.
3.
Select the desired parameter to change.
4.
Press the selection button for that parameter to toggle the available selections.
5.
Repeat steps 2 to 4 to program each AEC channel.
6.
When finished the AEC SETUP, press EXIT to return to the GEN
CONFIGURATION menu.
3C.5.6 AEC Calibration
Refer to chapter 3 section 3D, AEC CALIBRATION.