3D
AEC Calibration
CPI Canada Inc.
Page 3D-12
Rev. D Millenia / Indico 100 Series Service Manual Ch # 740904-06
3D.2.4 AEC
Pickup
Connections
(Overview)
Review the applicable sections of chapter 3 section B and C for interfacing AEC devices and programming
the image receptors to select the correct AEC device (or no AEC if desired). It should be ensured that each
receptor used for AEC has a bucky or equivalent. A typical R&F room configuration will consist of the
following:
•
Table bucky
•
Wall bucky
•
Spot film bucky
•
Aux - digital acquisition
Refer to Figure 3D-7 for typical AEC connections. This is a simplified view only, refer to figures 3D-8 to 3D-11
for AEC board layouts used in Millenia and Indico 100 generators. Refer to chapter 1E for the AEC board
location in your generator.
Figure 3D-7: AEC pickup connections