CPI Canada Inc
Troubleshooting
5
Millenia / Indico 100 Series Service Manual Ch # 740904-09 Rev. B
Page 5-7
5.2.3 Error
Messages
(Cont)
ERROR CODE
MESSAGE
PROBLEM
ACTION
E048 FOCUS
MISMATCH
Focus
selected
does
not
match current focus enabled
by power supply.
Check power supply
interface cables
between power
supply and generator
CPU board.
E049
NOT ENABLED
Requested function not
programmed to be enabled.
Reprogram to enable
function.
E050
GEN DATA ERROR
Generator detects corrupt
generator limit data.
Reprogram generator
limit data or set
factory defaults.
E051
AEC DEVICE ERR
Generator has detected no
feedback signal from AEC
device.
1. Check that X-ray
tube is pointing at
correct AEC
device.
2. Check AEC cable
for damage and
proper
connection.
E052 HIGH
SF
CURRENT
Generator
detects
small
focus filament current
greater than limits in standby
mode.
Check small focus
filament board.
E053 HIGH
LF
CURRENT
Generator
detects
large
focus filament current
greater than limits in standby
mode.
Check large focus
filament board.
E054
AEC OUT OF RANGE
AEC reference has reached
a maximum or minimum
limit.
Re-adjust AEC
calibration including
density to operate
within AEC range (O
to 1OVDC).
E055
NO FIELDS ACTIVE
AEC enabled but no fields
are selected.
Select AEC field(s).
E056 NO
TUBE
SELECTED
All
Receptors
have
no
X-ray
tube programmed.
Program receptor(s)
with tube number.
E057
AEC STOP ERROR
AEC stop signal (P.T. stop
signal) is active low
indicating exposure is
finished during prep state.
1. Check that P.T.
ramp does not
exceed P.T.
reference during
prep state.
2. Check AEC device
for proper
operation.