CPI Canada Inc.
Installation
2
Indico 100 Series Service Manual Ch # 740895-04 Rev. H
Page 2-35
2.19.1 Entering Into Programming/Calibration Mode (Cont)
4.
Enter the factory default password by pressing the following button sequence: [1] - [2] - [3] - [4].
5.
The APR menu will now display the following:
* GENERATOR SETUP *
UTILITY
APR EDITOR: DISABLED
GEN CONFIGURATION
DATA LINK
EXIT SETUP
2.19.2 Menu
Selections
UTILITY
Allows access to the following functions:
SET TIME & DATE:
Allows setting of: YEAR, MONTH, DAY, HOUR, and MINUTE
ERROR LOG:
Allows review of the generators error log for recorded errors.
STATISTICS:
Allows review of the tube exposure counter(s), the fluoro exposure counter if applicable, and the generator
accumulated exposure counter. Also allows resetting of the tube exposure counter(s), and the fluoro
exposure counter if used.
CONSOLE:
Allows setting the following parameters for the installation: key speed (scroll rate for the displays), speaker
volume, LCD screen mode, and APR mode.
APR EDITOR
This enables or disables technique factor changes. When APR EDITOR is set to ENABLED, the
technique factors may be changed. The new values may then be saved to overwrite the existing values.
Reference chapter 3C for further information.