CPI Canada Inc
Troubleshooting
5
Millenia / Indico 100 Series Service Manual Ch # 740904-09 Rev. B
Page 5-9
5.2.3 Error
Messages
(Cont)
ERROR CODE
MESSAGE
PROBLEM
ACTION
E101
CAL_DATA LIMIT
Auto calibration has
exceeded data table length
due to an excessive number
of exposures.
1. Check to see if
the filament
standby current is
too low.
2. Retry auto
calibration.
E102
CAL_MAX FIL ERR
Maximum filament current
for the selected focus has
been exceeded.
1. Check to see if
the maximum
filament current
limit can be
increased.
2. Retry auto
calibration.
E103 CAL_MAN
TERM Operator
released
exposure
button during auto
calibration.
Retry auto
Calibration.
E104 CAL_NO
MA
No
mA
feedback
detected
during auto calibration.
Check power supply
Interface cables
between HF power
supply and generator
CPU board.
E105 CAL_MIN
MA
ERR
Minimum
generator
mA
was
exceeded at start of
calibration. This is usually
caused by too high a
filament standby current on
the primary and or
secondary filament. (Primary
is the current filament being
calibrated, secondary is the
other filament and applies
only to generators with two
independent filament
supplies).
Reduce filament
standby current on
primary and/or
secondary filament.