CPI Canada Inc.
Programming the Generator
3C
Millenia / Indico 100 Series Service Manual Ch # 740904-05 Rev. B
Page 3C-11
3C.5.1 Tube Selection (cont)
5.
Select the second X-ray tube if desired by repeating the previous steps (two
tube generators only).
PLEASE ENSURE THAT THE SELECTED X-RAY TUBE STATOR(S) ARE COMPATIBLE
WITH THE LOW SPEED OR DUAL SPEED STARTER IN YOUR GENERATOR.
6.
Select EXIT to return to the GEN CONFIGURATION menu (two tube
generators only).
3C.5.2 Generator Limits
The GENERATOR LIMITS menu allows the setting of generator limits as defined below.
* GENERATOR LIMITS*
MAX KW: 80
MAX MA: 1000
MIN MA: 10
+
MAX MAS 630
-
EXIT
Definitions of generator limits as used in this section.
•
MAX KW
Sets the maximum generator kW limit.
•
MAX MA
Sets the maximum generator mA limit.
•
MIN MA
Sets the minimum generator mA limit.
•
MAX MAS
Sets the maximum generator mAs limit.
BEFORE MAKING ANY CHANGES IN THIS SECTION, PLEASE CONSULT THE X-RAY TUBE DATA
SHEETS TO ENSURE THAT THE PROPOSED CHANGES DO NOT EXCEED THE MANUFACTURERS
RECOMMENDED LIMITS.
NOTE:
BEFORE CHANGING GENERATOR LIMITS, PLEASE FILL IN THE X-RAY
TUBE AND GENERATOR PARAMETER WORKSHEET. A BLANK FORM
THAT SHOULD BE PHOTOCOPIED IS LOCATED IN SECTION 3A 3.0. THIS
ALLOWS RECORDING OF THE DEFAULT VALUES AND THE NEW
(CHANGED) VALUES.
Use these steps to set the generator limits.
Step
Action
1.
From the GEN CONFIGURATION menu select GENERATOR LIMITS.
2.
Select a limit to be changed. Refer to the previous definitions.
3.
Use the + and - buttons to change the selected values.
4.
When finished altering the generator default values, press EXIT to return to the
GEN CONFIGURATION menu.