CPI Canada Inc.
System Interfacing
3B
Indico 100 Series Service Manual Ch # 740895-05 Rev. E
Page 3B-5
3B.2.9 Room Interface Board
All the necessary inputs and outputs for the external room equipment are located on this board.
•
TB1 - Bucky 3 and tomo bucky select, spare input,
•
TB2 - Bucky 1 and 2 select, collimator and bucky inputs.
•
TB3 - Tomographic inputs and outputs, collimator bypass.
•
TB4 - Room light, thermal switches, door interlock, tube indicators.
•
TB5 - Multiple spot input, I.I. mag. select.
•
TB6 - Remote prep, exposure and fluoro inputs, I.I. safety input, ALE output, spare output.
3B.3.0 FEATURES OF THE ROOM INTERFACE BOARD
Refer to the following schematics and figures.
•
Figure 3B-2, room interface board layout.
•
Generator interface board schematic, drawing 732175 (sheet 1).
•
Room interface board schematic, drawing 733182.
LIVE C O N TAC T
D RY C O N TAC T
LIVE C O N TAC T
D RY C O N TAC T
D RY C O N TA CT
24 VD C
110 AC
22 0AC
J W 1-JW 5
24 V D C
J W 6-JW 8
LIV E C ON TA CT
D RY C O N TA CT
T B 1- TB 6
Figure 3B-2: Room interface board