Technology > Embedded RDS > E-RDS Setup > Initializing E-RDS settings
Technology > Embedded RDS > E-RDS Setup > Initializing E-RDS settings
b. Select #E-RDS SWITCH
、
and set value "
1
" then touch [OK]. The communication with
UGW is enabled by the above setting.
0 : OFF( initial value): e-Maintenance/ imageWARE Remote is disabled.
1 : ON: e-Maintenance/ imageWARE Remote is enabled
The following settings i.e. RGW-PORT and RGW-ADR in servicemode must not be
change unless there are specific instructions to do so. Changing these values will
cause error in communication with UGW.
c. Select COM-TEST and then touch [OK]. This initiates the communication test between
the device and the UGW. If the communication is successful, OK is displayed. If NG (failed)
appears, refer to the troubleshooting guide and repeat until OK! is displayed.
MEMO:
The communication results with UGW can be distinguished by referring to the
communication log. By performing the communication test with UGW, E-RDS acquires
schedule information and starts monitoring and meter reads operation.
Communication log
(1) COM-LOG
#E-RDS COM-LOG
Move on to COM-TEST by using Left/ Right allow key.
The maximum 5 records logged
Only for the head portion of the error details written on COM-LOG is displayed.
(2) Details of communication log
You can use communication log to verify that a detailed error code SOAP method name,
etc. are displayed at the time of communication test failure.
By pressing ? key or OK key, progressive information (generating date/ time of error) is
displayed.
Error details will be displayed when OK key is again pressed in the state line where
generating time is displayed.
Error information: within maximum 128 character strings (NULL is not included).
Refer to the error list for the message strings displayed on the COM-LOG.
•
•
■
•
•
•
Initializing E-RDS settings
It is possible to return E-RDS Settings to factory-shipments value.
1)
Select #CLEAR ERDS-DAT by using Left/ Right allow key.
2)
When [OK] is touched, a confirmation message "OK?" is displayed.
3)
If agree upon initialization, following E-RDS settings, internal data, and Alarm filtering
information will be purged.
#E-RDS > E-RDS SWITCH
#E-RDS > RGW-ADDRESS
#E-RDS > RGW-PORT
#E-RDS > COM-LOG
In case of replacing the CA certificate file, even #CLEAR is executed, the status is not
returned to the factory default.
When installing the certificate file other than the factory default CA certificate file, it is
required to delete the certificate file after E-RDS initialization and install the factory
default CA certificate file.
■
•
•
•
•
Содержание IMAGERUNNER 2530
Страница 1: ...9 8 7 6 5 4 3 2 1 imageRUNNER 2530 2525 2520 Series Service Manual...
Страница 4: ...Blank Page...
Страница 16: ...1 1 Product Overview Product Overview Product Lineup Feature Specifications Name of Parts...
Страница 111: ...3 3 Periodical Service Periodical Service Consumable Parts and Cleaning Parts...
Страница 159: ...5 5 Adjustment Adjustment Outline Adjustment when replacing parts Image position adjustment...
Страница 166: ...6 6 Troubleshooting Troubleshooting Upgrading Targets and Procedure...
Страница 171: ...7 7 Error Code Error Code Overview Error Code Jam Code Alarm Code...
Страница 186: ...8 8 Service Mode Service Mode Outline Details of Service Mode...
Страница 321: ...Service Tools General Circuit Diagram Appendix...