56 | Chapter 2: Installation
Install the Dispense Cassette
Prerequisites:
o
Review the
Dispense Cassette Diagram on page 54
to learn the names of the
components.
o
Move the
Pump Cover
away from the pump to its
OFF
position.
o
When you have two Peri-pumps installed, put the Primary pump's Tip Holder in the
left bracket and the Secondary pump's Tip Holder in the right bracket.
o
Release the pump’s stainless steel plate: Use your right hand to release the
spring-loaded cassette latch (on the right side of the caution symbol on the pump)
and use your left hand to lift the stainless steel plate up and out.
The BioTek logo on the Tip Holder faces the Peri-pump. The logo on the Center Holder
faces down, away from the pump.
Note:
See Install RAD Technology Cassette on page 58
for single-tube and bulk
cassette instructions.
MultiFlo™ FX Multi-Mode Dispenser