176 | Chapter 4: Maintenance
Recommended Maintenance Schedule
Tasks
Frequency
Daily
Overnight/
Multi-Day
Weekly
Periodic/
Monthly
Before
storage/
shipment
Washer
Run W-DAY_RINSE
ü
ü
Run AutoPrime
ü
Run W-OVERNIGHT_LOOP
ü
Run W-RINSE_AND_SOAK
ü
Peri-pump
ü
ü
Record number of plates
processed with cassette
ü
Syringe Dispenser
Run S-DAY_RIN SE_A&B
ü
ü
Run AutoPrime
ü
Components
Remove protein residuals and
fungi growth, (if necessary)
ü
ü
ü
Check/empty waste bottles
ü
ü
Clean bottles
ü
ü
Clean plate carrier system
ü
ü
Clean washer manifold
ü
ü
Clean aspirate and dispense tubes
ü
ü
Clean exterior surfaces
ü
Clean Syringe dispenser manifold
ü
ü
Clean Syringe dispenser tubes
ü
ü
MultiFlo™ FX Multi-Mode Dispenser