Biomagnetic Separation - Magnetic Bead Assays | 149
96 Ring Magnet
PN: 7102216
This magnetic bead separator uses
329 of VP’s 52 MGO magnets (7094
Gauss). The magnets are arranged
around each well, pulling the
magnetic beads to the bottoms and
edges of the wells. Aspirate from
the center of the well (the default
position), when using this magnet.
384 Ring Magnet
PN: 7102215
This magnetic bead separator uses
425 of VP’s 52 MGO magnets (6994
Gauss). The magnets are aligned
with the intersections of the wells,
pulling the magnetic beads to the
bottoms and edges of the wells.
Every well is circled by 4 magnets.
Aspirate from the center of the well
(the default position), when using this
magnet.
How to determine the Magnet Adapter Height Offset
Use calipers for the best results or another measuring device.
1. Put a magnet in the adapter, and the
type of plate you will be using in
place on top of it.
2. Measure the distance from the
bottom of the adapter to the top of
the plate in millimeters (mm).
3. Remove the plate, and measure it
alone.
4. Subtract the stand-alone plate
measurement from the measurement
value of the combined elements.
This is the value to use.
Examples
In our testing we found these offset values to be effective:
BioTek Instruments, Inc.
Содержание MultiFlo FX
Страница 1: ...Operator s Manual Microplate Dispenser MultiFlo FX Microplate Dispenser ...
Страница 2: ......
Страница 67: ...Set Up the Peri pump Dispenser 45 MultiFlo FX with closed Peri pump cover panel BioTek Instruments Inc ...
Страница 236: ...214 Chapter 4 Maintenance MultiFlo FX Multi Mode Dispenser ...
Страница 280: ...258 Chapter 5 Qualification MultiFlo FX Multi Mode Dispenser ...
Страница 292: ......