222 | Chapter 5: Qualification
Tests – Solutions
Cassette Types
1 µL
1 µL 1536
5 µL
10 µL
1 µL
√
√
5 µL
√
10 µL
√
√
50 µL
√
100 µL
√
1536
√
8-to-1
√
l
1 µL Test
: Confirms the performance of the 1
µ
L cassettes when dispensing a single
aliquot (1/4 turn of pump) into each well of the plate. It dispenses 1
µ
L into each well
using the
1 µL Solution
, and requires an additional 150 µL of deionized or distilled
water to raise the fluid level for optimal reading.
n
A single aliquot for a cassette type is the smallest volume unit recommended
for it. 1 µL for the 1 µL cassette, 5 µL for the 5 µL cassette, and 10 µL for the
10 µL cassette (except that later model instruments can dispense 0.5 µL/well
using a 1 µL cassette).
l
5 µL Test
: Confirms the performance of the 5
µ
L cassettes when dispensing a single
aliquot (1/4 turn of pump) into each well of the plate. It dispenses 5
µ
L into each well
using the
5 µL Solution
, and requires an additional 150 µL of deionized or distilled
water to raise the fluid level for optimal reading.
l
10 µL Test
: Confirms the performance of the 1
µ
L cassettes when dispensing 10 aliquots
(2 1/2 turns of pump) and the 10
µ
L cassettes when dispensing a single aliquot (1/4 turn
of the pump) into each well of the plate. It dispenses 10
µ
L into each well using the
10
µL Solution
, and requires an additional 100 µL of deionized or distilled water to raise
the fluid level for optimal reading.
l
50 µL Test
: Confirms the performance of the 5
µ
L cassettes when dispensing 10 aliquots
(2 1/2 turns of pump) into each well of the plate. It dispenses 50
µ
L into each well using
the
50 µL solution
, and requires an additional 100 µL of deionized or distilled water to
raise the fluid level for optimal reading.
l
100 µL Test
: Confirms the performance of the 10
µ
L cassettes when dispensing 10
aliquots (2 1/2 turns of pump) into each well of the plate. It dispenses 100
µ
L into each
well using the solution called
100 µL solution
, and requires an additional 50 µL of
deionized or distilled water to raise the fluid level for optimal reading.
l
1536 Test
: Confirms the alignment of the tips; that the cassette is firing straight into
the wells. Dispenses 6 µL into columns 2, 4, 19-30, 45, 47 of a 1536-well plate using the
"1536 solution." Also requires performing the 1 µL Test described above.
MultiFlo™ FX Multi-Mode Dispenser
Содержание MultiFlo FX
Страница 1: ...Operator s Manual Microplate Dispenser MultiFlo FX Microplate Dispenser ...
Страница 2: ......
Страница 67: ...Set Up the Peri pump Dispenser 45 MultiFlo FX with closed Peri pump cover panel BioTek Instruments Inc ...
Страница 236: ...214 Chapter 4 Maintenance MultiFlo FX Multi Mode Dispenser ...
Страница 280: ...258 Chapter 5 Qualification MultiFlo FX Multi Mode Dispenser ...
Страница 292: ......