114 | Chapter 3: Operation
Minimum Prime Volumes for the Strip Washer
Flow rate:
Wash Manifold Type
6-Well
12-Well
24-Well
48-Well
96/384-Well
1
310
315
320
360
320
2
130
180
160
360
160
3
80
180
80
360
80
4
240
180
240
360
240
5
240
240
160
360
160
6
300
240
320
360
320
7
440
300
440
420
440
8
400
420
400
420
400
9
400
480
480
600
480
10
800
780
800
630
800
11
880
870
880
900
880
Bottom Wash
Bottom washing adds an initial dispense/aspirate sequence to the protocol. Fluid
is simultaneously dispensed and aspirated to create cleaning turbulence (at the
specified height). The manifold descends to aspirate again and ends with a final
dispense to fill the wells.
Designed for strongly bound molecules in assays that require vigorous washing,
e.g. ELISA, adding a bottom wash to the protocol has shown reduced background
noise without increasing the number of wash cycles.
Adjust the Z-axis setting to position the manifold deep in the well to target the
solid phase of coated assay plates.
Bottom Wash Volume:
Enter the volume of wash solution to dispense per well
during the bottom wash.
Flow Rate:
Select a rate between 3-11 for best results. Cell wash options are
designed for gentle washing, while bottom wash is designed for vigorous washing.
MultiFlo™ FX Multi-Mode Dispenser
Содержание MultiFlo FX
Страница 1: ...Operator s Manual Microplate Dispenser MultiFlo FX Microplate Dispenser ...
Страница 2: ......
Страница 67: ...Set Up the Peri pump Dispenser 45 MultiFlo FX with closed Peri pump cover panel BioTek Instruments Inc ...
Страница 236: ...214 Chapter 4 Maintenance MultiFlo FX Multi Mode Dispenser ...
Страница 280: ...258 Chapter 5 Qualification MultiFlo FX Multi Mode Dispenser ...
Страница 292: ......