MultiFlo FX Dispenser Comparison | 7
MultiFlo FX Dispenser Comparison
Counting the washer as a potential dispenser, the MultiFlo FX offers three distinct
dispensers. Here is a comparison of the devices:
n
For
precious reagents
use the Peri-pump to preserve unused fluids. It has
the shortest, most visible fluid path, and a Purge capability to reverse the fluid
flow to recover fluid from the tubing. Another advantage is the ability to
dedicate a dispense cassette to use with one reagent only, reducing the
amount of priming required prior to use. And, you can shorten the tubing to
further reduce the dead volume:
See Shorten the Dispense Cassette
n
offers the best tools for preserving cells and reagents.
n
All MultiFlo FX devices are capable of dispensing up to 30,000 µL. BioTek's
recommended maximum dispense volumes are cited here.
Device
Volume
range
µL/well
Precision
Accuracy
Approximate Dead
volume
Peri-pump
0.5‡, and
1-3000*
<10% CV
@ 1
µL/well
(typical
<3% CV)
+/-10%
(typical
+/-3%) @
1 µL/well
Cassette Type:
1 µL
1.20 mL
5 µL
4.23 mL
10 µL
7.36 mL
RAD
Technology
0.5‡, and
1-
30,000*
<10% CV
@ 1
µL/well
(typical
<3% CV)
+/-10%
(typical
+/-3%) @
1 µL/well
Type
Dead Vol.
1 µL
150 µL
5 µL
530 µL
10 µL
920 µL
Syringe 8-
tube/6-well
10-3000
<5% CV @
20 µL/well
±2 µL @
10
µL/well
For all manifold types:
12 mL
Syringe 16-tube
5-3000
<10% CV
@ 5
µL/well
±2 µL @
10
µL/well
Syringe 12-/24-
5-30,000
<3% CV @
±2 µL @
BioTek Instruments, Inc.
Содержание MultiFlo FX
Страница 1: ...Operator s Manual Microplate Dispenser MultiFlo FX Microplate Dispenser ...
Страница 2: ......
Страница 67: ...Set Up the Peri pump Dispenser 45 MultiFlo FX with closed Peri pump cover panel BioTek Instruments Inc ...
Страница 236: ...214 Chapter 4 Maintenance MultiFlo FX Multi Mode Dispenser ...
Страница 280: ...258 Chapter 5 Qualification MultiFlo FX Multi Mode Dispenser ...
Страница 292: ......