218 | Chapter 5: Qualification
System Self-Test, Verify Information
Perform these steps to verify software information and run a system self-check:
Prerequisite: When controlling the instrument with the LHC, ensure that it is
attached to the host computer and turned on, and then launch the LHC software.
To run the System Self-Test:
LHC
Touch screen
Select
Tools>Instrument
Utilities>General Settings
.
Under
Instrument Functions
,
click Perform
Self-check
.
1. Select
Instrument
at the
Home
screen.
2. On the
Other
screen, press
Run Self-
Check
.
Test Results:
l
If the test fails, an error code displays. If this happens, find the error code in
the MultiFlo FX Operator’s Manual to determine its cause. If the problem is
something you can fix, turn off the instrument, fix the problem, and then turn
the instrument back on and re-run the test. If the problem is not something
you can fix, contact BioTek’s Technical Assistance Center.
Record Onboard Software
Record the software part number and version installed on the MultiFlo FX.
LHC
Touch screen
1. Select
Tools>Instrument
Utilities
2. Select the
Software
tab and from
the Basecode Software
Information section record the
l
Software Version and part
number
1. Select
Instrument
at the
Home
screen.
2. Record the
Software
part number
and version from the
About
screen.
MultiFlo™ FX Multi-Mode Dispenser
Содержание MultiFlo FX
Страница 1: ...Operator s Manual Microplate Dispenser MultiFlo FX Microplate Dispenser ...
Страница 2: ......
Страница 67: ...Set Up the Peri pump Dispenser 45 MultiFlo FX with closed Peri pump cover panel BioTek Instruments Inc ...
Страница 236: ...214 Chapter 4 Maintenance MultiFlo FX Multi Mode Dispenser ...
Страница 280: ...258 Chapter 5 Qualification MultiFlo FX Multi Mode Dispenser ...
Страница 292: ......