Peri-pump Dispense Precision & Accuracy Test Worksheet
8-to-1 RAD Cassette Test
100 µL Dispense Precision Test
Standard Deviation (SD):
Mean Absorbance (sum of all wells ÷ 24)
% CV (SD ÷ Mean x 100)
%
% CV must be < 2.5%
c
Pass
c
Fail
100 µL Dispense Accuracy Test
Total Dispense (Actual) Weight:
grams
Expected Weight:
(mL/well x number of wells dispensed)
grams
% Accuracy Error:
(Actual Weight – Expected Weight) ÷ Expected Weight x 100
%
% Accuracy Error must be < 4.0%
c
Pass
c
Fail
Visual verification that no well varies
considerably from the others
c
Pass
c
Fail
Cassette Serial Num-
ber:
Tests Performed By:
Date:
Reviewed/Approved
By:
Date:
MultiFlo FX Qualification Procedures
© BioTek® Instruments, Inc.
Содержание MultiFlo FX
Страница 1: ...Operator s Manual Microplate Dispenser MultiFlo FX Microplate Dispenser ...
Страница 2: ......
Страница 67: ...Set Up the Peri pump Dispenser 45 MultiFlo FX with closed Peri pump cover panel BioTek Instruments Inc ...
Страница 236: ...214 Chapter 4 Maintenance MultiFlo FX Multi Mode Dispenser ...
Страница 280: ...258 Chapter 5 Qualification MultiFlo FX Multi Mode Dispenser ...
Страница 292: ......