Hazards and Precautions| xv
Contact BioTek for more information and prior to using other questionable
chemicals.
Caution: Bovine Serum Albumin.
Solutions containing proteins, such as
bovine serum albumin (BSA), will compromise the instrument’s performance
over time unless a strict maintenance protocol is adhered to. See
Maintenance
procedures regarding BSA.
Caution: Power Supply.
Only use the power supply shipped with the
instrument. Operate this power supply within the range of line voltages listed
on it.
Caution: Disposal.
This instrument contains printed circuit boards and wiring
with lead solder. Dispose of the instrument according to Directive 2002/96/EC,
“on waste electrical and electronic equipment (WEEE),” or local ordinances.
Caution: Warranty.
Failure to follow preventive maintenance protocols may
void the warranty.
Caution: Shipping Hardware.
All shipping hardware (e.g., shipping bracket
etc.) must be removed before operating the instrument and reinstalled before
repackaging the instrument for shipment.
Caution:
Do not run the Peri-pump without a cassette installed on the pump.
Caution: Electromagnetic Environment.
Per IEC 61326-2-6 it is the user’s
responsibility to ensure that a compatible electromagnetic environment for this
instrument is provided and maintained in order that the device will perform as
intended.
Caution: Electromagnetic Compatibility.
Do not use this device in close
proximity to sources of strong electromagnetic radiation (e.g., unshielded
intentional RF sources), because these may interfere with the proper operation.
BioTek Instruments, Inc.
Содержание MultiFlo FX
Страница 1: ...Operator s Manual Microplate Dispenser MultiFlo FX Microplate Dispenser ...
Страница 2: ......
Страница 67: ...Set Up the Peri pump Dispenser 45 MultiFlo FX with closed Peri pump cover panel BioTek Instruments Inc ...
Страница 236: ...214 Chapter 4 Maintenance MultiFlo FX Multi Mode Dispenser ...
Страница 280: ...258 Chapter 5 Qualification MultiFlo FX Multi Mode Dispenser ...
Страница 292: ......