235 | Chapter 5: Qualification
Syringe Dispenser Liquid Tests
Dispense Precision and Accuracy Specifications
n
Important
: For
IQ/OQ/PQ
testing purposes 1.0% tolerance has been added
to some of the
for Precision %CV to accommodate
variations in test solutions, off-peak wavelengths, reader errors, and pipette
errors. % Accuracy Error is calculated for the dispense volume specified in the
respective test procedure in adherence to the published specifications.
Test #
Plate Type-Manifold
Precision
Accuracy
2
384-well plate
< 3% CV @ 80 µL/well
< 1.25% @ 80 µL/well
1
16-tube manifold
< 6% CV @ 20 µL/well
< 5% @ 20 µL/well
2
96-well plate
< 3% CV @ 160 µL/well
< 1.25% @ 160 µL/well
1
8- & 16-tube manifold
< 6% CV @ 40 µL/well
< 5% @ 40 µL/well
3
1536-well plate
< 12% CV @ 6 µL/ well
< 5% @ 6 µL/well
32-tube manifold
Low-Density-Plate-Specific Manifolds
Dispense Precision
Refer to the values below for dispense precision when measured in full plates with the
specified volume of deionized water with 0.1% Tween 20 and FD&C#1 blue dye solution
and read at 630 nm/405 nm.
Plate Type
Performance
Volume (µL/well)
6-Well
≤6.0% CV
5560
12-Well
≤4.0% CV
2240
24-Well
≤4.0% CV
1120
48-Well
≤4.0% CV
560
Dispense Accuracy
is measured gravimetrically when dispensing any volume of room-
temperature deionized water using
Flow Rate
2. For all plate-specific manifolds, dispense
accuracy shall be 2 µL or 1%, whichever is greater.
MultiFlo™ FX Multi-Mode Dispenser
Содержание MultiFlo FX
Страница 1: ...Operator s Manual Microplate Dispenser MultiFlo FX Microplate Dispenser ...
Страница 2: ......
Страница 67: ...Set Up the Peri pump Dispenser 45 MultiFlo FX with closed Peri pump cover panel BioTek Instruments Inc ...
Страница 236: ...214 Chapter 4 Maintenance MultiFlo FX Multi Mode Dispenser ...
Страница 280: ...258 Chapter 5 Qualification MultiFlo FX Multi Mode Dispenser ...
Страница 292: ......