Protocol Parameters Tables | 105
Step
Option
Description/Values range
Default
values
d?:
Plate
type:
Select the plate type, and optionally, limit the
columns dispensed to. For high density plates,
skipping certain "Rows" sections is supported, see
Channel Dispenser Dispense Pattern on page 139
96
Position:
X- and Y- horizontal axes, Z- height (vertical) axis
can be adjusted to improve performance. X- and Y-
axes default to 0 steps for all plate types.
Z = plate
type
dependen
t
Pre-
dispense:
Also called "tip priming," pre-dispense normalizes the
tips to ensure precise fluid distribution. Pre-dispense
is recommended for most applications. Set volume
and number of pre-dispenses, except, when
dispensing 0.5 µL, the volume is preset: 0.5 µL for 4
cycles by default.
10 µL
2 cycles
Purge
To preserve fluid in the tubing by pumping it back into
the supply vessel, i.e. reverses the flow direction.
Volume:
1-3000 µL
300
Duration:
1-300 seconds
3
Note:
*Half-microliter (0.5 µL) dispensing requires a 1 µL cassette. The Cassette
Requirement Mode behavior is implemented whenever a 0.5 µL volume is requested.
Make sure a 1 µL cassette is installed and the Cassette Type setting matches it.
Only a 0.5 µL dispense volume can be requested, not 1.5 or 2.5 µL, for example. Perform
two dispense steps, one for the half microliter and another at a full increment to achieve
these volumes.
Important:
Remember that the Peri-pump, the Syringe dispenser and Washer manifolds
travel on the same dispense arm. Keep this in mind when adjusting the Z-axis or
Dispense Height of any dispenser. A too low setting can cause a dispenser to crash into
the plate.
Uninstall the dispense cassette, if it is not required by a protocol, to prevent collisions
when trying to position the Syringe or Wash dispenser as close as possible to the wells.
And/or, consider increasing the
Plate Clearance Height
setting to increase the travel
height.
BioTek Instruments, Inc.
Содержание MultiFlo FX
Страница 1: ...Operator s Manual Microplate Dispenser MultiFlo FX Microplate Dispenser ...
Страница 2: ......
Страница 67: ...Set Up the Peri pump Dispenser 45 MultiFlo FX with closed Peri pump cover panel BioTek Instruments Inc ...
Страница 236: ...214 Chapter 4 Maintenance MultiFlo FX Multi Mode Dispenser ...
Страница 280: ...258 Chapter 5 Qualification MultiFlo FX Multi Mode Dispenser ...
Страница 292: ......