Install RAD Technology Components | 57
1. Slide the
Tip Holder
into the
dispense arm. The Tip Holder’s
front plate with the BioTek logo
or steel plate faces the pump.
Make sure the tip holder is level
and snapped into place.
l
Two Peri-pumps: use left bracket for
Primary, right bracket for
Secondary.
l
RAD technology
exception: attach
the tip holder to the dedicated
.
2. Extend the rest of the cassette
under the Peri-pump, so you
can slide the
Center Holder
into
its slot on the underside of the
pump. A tab on the back of the
center holder fits into a notch
on the pump. The label on the
center holder faces down.
3. Align the
Tube Tensioner
with
the stainless steel plate as it
wraps around and up against
the pump. Be sure the knobs on
top of the tensioner fit correctly
into the grooves in the stainless
steel plate as you move both
parts up and around the pump
and click the steel plate into
place.
4. Return the Pump Cover to its
RUN
position covering the pump.
5. Lift the
Tube Organizer
over the pump cover. Place it in the fluid vessel, when
you’re ready.