126 | Chapter 3: Operation
1. Highlight a step or
the <end of steps>
placeholder and press
Add
and then
BioStack
. This adds a
start step, BioStack
delivers and an end
step, BioStack returns
plate. The end step is
highlighted.
2. Press the desired
action button and
define the parameters
of the steps to be
performed on each
plate.
Make sure the steps to be performed on the plates are inside the 'BioStack
delivers' and 'BioStack returns' steps.
3. Optionally, highlight the <end of steps> placeholder, press
Add
and add a
Restack
step to move the plates back to the input stack. A
Restack
step must
be outside the BioStack step container.
4. When all the protocol steps are defined, click
Validate
to make sure the
protocol is correctly designed.
5. Save the protocol.
6. At runtime, except when using plates with lids, you can define the
Number of
plates to process
in the
Run Info
section of the run screen: enter a specific
number or select the "Entire input stack" for processing.
How to run plates with lids
1. Select the
Info
button and fill the checkbox
to
Use Lids
.
2. Select a file in the
Lid Type
field that most
closely matches your plate type. (Touch
screen: press the field to select a file.) (Select
View definition
to see the file's details.)
Conversely, de-select
Use Lids
to process plates without them.
MultiFlo™ FX Multi-Mode Dispenser
Содержание MultiFlo FX
Страница 1: ...Operator s Manual Microplate Dispenser MultiFlo FX Microplate Dispenser ...
Страница 2: ......
Страница 67: ...Set Up the Peri pump Dispenser 45 MultiFlo FX with closed Peri pump cover panel BioTek Instruments Inc ...
Страница 236: ...214 Chapter 4 Maintenance MultiFlo FX Multi Mode Dispenser ...
Страница 280: ...258 Chapter 5 Qualification MultiFlo FX Multi Mode Dispenser ...
Страница 292: ......