112 | Chapter 3: Operation
Minimum Dispense Volumes (µL/well) per Manifold
Rate 96/384-well 6-well 12-well 24-well 48-well
1 CW
40
155
105
80
60
2
20
65
60
40
60
3
20
65
60
40
60
4
30
120
60
60
60
5
20
120
80
60
60
6
40
150
80
80
60
7
55
225
100
110
70
8
50
225
140
105
70
9
60
225
160
120
100
10
100
400
260
200
105
11
110
450
290
220
150
Aspirate Travel Rates
Rate
mm/sec
0 CW
1.0
for 6-, 12-, 24-well
plates
1
4.1 +1
Slowest
1 CW
4.1
2
5.0 +1
2 CW
5.0
3
7.3 +1
Default rate
3 CW
7.3
4
9.4 +1
4 CW
9.4
5
9.4 +2
Fastest
6 CW
14.7
Recommended rate
for Cell Wash
protocols
MultiFlo™ FX Multi-Mode Dispenser