196 | Chapter 4: Maintenance
RAD Single-tube Cassette - Unclog the Tip
Disassemble the cassette to unclog the tip:
Single-tube RAD Cassette
1. Use the screwdriver provided with the MultiFlo
FX to disassemble the cassette's tip holder.
Remove the two screws on the single-tube
cassette's cover to release its bottom plate.
2. Upturn the cassette's tip holder, and remove the
tubing.
3. Pull off the tip and follow the instructions to
flush the tip using the 10 cc syringe.
4. When finished, reassemble the tip holder:
l
Put the straight end of the tip into the bottom
of the tube leaving the tapered end exposed.
l
1 µL tubing: put the little block in the space
provided for it.
l
All tubing: slide the flared top of the tip into
its molded slot. Only the end of the tip
protrudes from the holder.
l
Replace the bottom plate: it only fits one way.
l
Holding the pieces together, turn it over and
insert the two screws. Tighten each screw half
way, rotating between the two before
tightening them all the way.