Install the Secondary Peri-Pump | 49
with the support bracket and the
four screws removed from the
instrument earlier. Partially tighten
the screws until all 4 are in place,
then tighten all the screws.
Hide the bracket's foot holds inside the case
when a Syringe dispenser will not be
installed on top of the second Peri-Pump.
Install the additional dispense cassette bracket:
This task is easier to perform when the cassette bracket is well lit. It's a tight space and
the second cassette bracket fits snugly onto the original bracket.
1. First, remove the four screws and
washers from the additional cassette
bracket using the small Allen
wrench supplied in the kit.
2. Slide the plate carrier to the left, out
of the way. Lower the dispense arm
to a workable position.
3. Use the text on the front of the
additional cassette bracket to orient
it correctly and place it around the
original bracket.
4. Use the four screws to secure the
bracket over the existing bracket.
Place and partially tighten the
screws until all four are in place,
then tighten all the screws.