74 | Chapter 2: Installation
Update the Instrument to use the Syringe Dispenser
An important part of installing the dual Syringe
Dispenser is updating the instrument's basecode
(internal software) with its calibration data. A data
sheet is affixed to the bottom of the Syringe pump
unit and to the front page of the installation
instructions.
Touch screen users: Use the USB flash drive and
enter both data sets, for the 8- and 16-Tube
Manifolds.
USB Flash Drive
: plug in the USB flash drive
containing Syringe calibration data to update the
MultiFlo FX automatically.
LHC
Touch screen
1. Connect the syringe pump unit
to the instrument with its serial
cable.
2. Turn on the instrument, launch
the LHC and make sure it is
communicating with the
MultiFlo FX (define the correct
COM port).
3. Select
Tools>Instrument
Utilities>Syringe Dispenser
.
4. Under
Syringe Dispenser
Assembly,
specify the type of
pump and the installed
manifold type and click
Send
to
download the data to the
instrument.
5. In the
Calibration Data
section,
enter the data points on the
label in the corresponding
fields, including the Serial
Number or plug the USB flash
1. Connect the syringe pump unit
to the instrument with its serial
cable.
2. Turn on the instrument, and
press Instrument.
3. Select Syringe and specify the
installed manifold type.
4. Under Calibration Data from,
press the link for each
manifold, one at a time, and
either Import from USB
or enter the Serial Number (SN)
and all the data points by
pressing each field to open the
number pad.
n
Enter both data sets.
5. Press Save this Data when
done and confirm that the
Checksum displayed onboard
matches the Checksum on the
MultiFlo™ FX Multi-Mode Dispenser
Содержание MultiFlo FX
Страница 1: ...Operator s Manual Microplate Dispenser MultiFlo FX Microplate Dispenser ...
Страница 2: ......
Страница 67: ...Set Up the Peri pump Dispenser 45 MultiFlo FX with closed Peri pump cover panel BioTek Instruments Inc ...
Страница 236: ...214 Chapter 4 Maintenance MultiFlo FX Multi Mode Dispenser ...
Страница 280: ...258 Chapter 5 Qualification MultiFlo FX Multi Mode Dispenser ...
Страница 292: ......