286 | Appendix A: Error Codes
Code
Message
What to do
309
RAD Y-axis motor interlock
safety switch open
Service Only.
Contact BioTek TAC.
400
Carrier X motor failed positional
verify
Run self test. If error reoccurs, contact
BioTek TAC.
401
Carrier Y motor failed positional
verify
Run self test. If error reoccurs, contact
BioTek TAC.
402
Dispense head motor failed
positional verify
Run self test. If error reoccurs, contact
BioTek TAC.
403
Secondary Peri-pump motor
failed positional verify
Service Only.
Contact BioTek TAC.
404
Syringe A motor failed
positional verify
Service Only.
Contact BioTek TAC.
405
Syringe B motor failed
positional verify
Service Only.
Contact BioTek TAC.
406
Peri-pump motor failed
positional verify
Service Only.
Contact BioTek TAC.
407
Strip washer syringe motor
failed positional verify
Service Only.
Contact BioTek TAC.
408
Aspirate head motor failed
positional verify
Service Only.
Contact BioTek TAC.
409
RAD Y-axis motor failed
positional verify
Service Only.
Contact BioTek TAC.
600-609
Specified motor currently in use
Service Only.
Contact BioTek TAC.
700 -
70F
Invalid motor number specified
Service Only.
Contact BioTek TAC.
900
Calibration failed. The
measured or calculated autocal
value is out of tolerance
Service Only.
Contact BioTek TAC.
A00
Invalid plate type selected
The currently selected plate type is not
supported with the currently installed or
requested hardware. If this is not the case,
contact BioTek TAC.
MultiFlo™ FX Multi-Mode Dispenser
Содержание MultiFlo FX
Страница 1: ...Operator s Manual Microplate Dispenser MultiFlo FX Microplate Dispenser ...
Страница 2: ......
Страница 67: ...Set Up the Peri pump Dispenser 45 MultiFlo FX with closed Peri pump cover panel BioTek Instruments Inc ...
Страница 236: ...214 Chapter 4 Maintenance MultiFlo FX Multi Mode Dispenser ...
Страница 280: ...258 Chapter 5 Qualification MultiFlo FX Multi Mode Dispenser ...
Страница 292: ......