288 | Appendix A: Error Codes
Code
Message
What to do
1301
Syringe module not connected
Make sure the Syringe module is correctly
connected using the new BioTek-provided
serial cable.
1302
Syringe initialization error
1303
Syringe sensor not cleared
error
1304
Invalid syringe dispense volume
See error 1308 below.
1305
Invalid syringe operation
Contact BioTek TAC.
1306
FMEA error on syringe A
Contact BioTek TAC.
1307
FMEA error on syringe B
Contact BioTek TAC.
1308
Invalid Syringe pre-dispense
volume
Protocol may have been written for a
different type of dispense manifold. Make
sure the Instrument Settings represent the
installed hardware. Modify the protocol to
match.
1309
Invalid Syringe prime volume
1310
Invalid Syringe manifold
1350
Peri-pump (PP) invalid dispense
volume
Contact BioTek TAC.
1351
PP invalid cassette type
Change the cassette type to match the
protocol requirement and rerun the protocol.
1352
PP invalid pre-dispense volume
Contact BioTek TAC.
1353
Multiple required cassettes
Make sure every Peri-pump step in the
protocol calls for the same cassette type. A
conflict was found.
1354
PP pump cover (safety door)
open
Close the pump cover door and rerun the
protocol.
1355
PP not installed
If you are trying to run a Secondary Peri-
pump, check the cabling in the back of the
instrument. Otherwise Contact BioTek TAC.
1356
PP Invalid dispense position
Contact BioTek TAC.
1357
PP two dispensers are expected
Contact BioTek TAC.
1358
PP Half UL not supported
0.5 µL Peri-pump dispensing is not
supported on this instrument.
MultiFlo™ FX Multi-Mode Dispenser
Содержание MultiFlo FX
Страница 1: ...Operator s Manual Microplate Dispenser MultiFlo FX Microplate Dispenser ...
Страница 2: ......
Страница 67: ...Set Up the Peri pump Dispenser 45 MultiFlo FX with closed Peri pump cover panel BioTek Instruments Inc ...
Страница 236: ...214 Chapter 4 Maintenance MultiFlo FX Multi Mode Dispenser ...
Страница 280: ...258 Chapter 5 Qualification MultiFlo FX Multi Mode Dispenser ...
Страница 292: ......