Perform the Syringe Dispense Precision & Accuracy Test | 238
manifolds:
Skip these protocol development steps for the 32-tube dispensers. And,
because you will save the protocols, you only need to create them one
time for the other manifolds.
1. Define a Dispense step for each test for each manifold, A and B:
Standard Syringe Manifolds
Manifold
Type
Plate
Type
Volume
(µL/wel
l)
Flow
rate
Pre-dispense (2 cyc)
Test 1:
16-Tube
384
20
2
20
8-Tube
96
40
2
40
Test 2:
16-Tube
384
80
2
80
8-Tube
96
160
2
160
Plate-Type Specific Manifolds
Parameters
6-Well
12-Well
24-Well
48-Well
Volume (µL/well)
5560
2240
1120
560
Flow Rate
5
5
5
5
Z-axis/Dispense Height 465
460
452
460
Pre-dispense (2 cycles)
320
220
160
320
2. Save the protocol.
4. Place a clean, empty microplate on the balance and tare the balance.
5. Place the microplate on the carrier and run the protocol (created in step 3 or
predefined).
6. Place the plate on the balance and record the
Total Dispense Weight
. This value
will be used to calculate the % Accuracy Error.
7.
For Test 1 Only
: Use a calibrated hand pipette or the Peri-pump to dispense
deionized water on top of the dye solution in the wells.
l
384-well:
Pipette 60
µ
L/well (resulting in 80
µ
L/well)
l
96-well:
Pipette 120
µ
L/well (resulting in 160
µ
L/well)
BioTek Instruments, Inc.
Содержание MultiFlo FX
Страница 1: ...Operator s Manual Microplate Dispenser MultiFlo FX Microplate Dispenser ...
Страница 2: ......
Страница 67: ...Set Up the Peri pump Dispenser 45 MultiFlo FX with closed Peri pump cover panel BioTek Instruments Inc ...
Страница 236: ...214 Chapter 4 Maintenance MultiFlo FX Multi Mode Dispenser ...
Страница 280: ...258 Chapter 5 Qualification MultiFlo FX Multi Mode Dispenser ...