Distillation Process Analyzer DPA-4.5
Installation and preparation for commissioning
6.2 Requirements for the installation location
Ambient conditions
The device is solely intended for use at a fixed location in an
analysis building that provides protection against the elements.
The following ambient conditions must be met:
Environment
Temperature range
5–40 °C
Relative humidity, non-corrosive
5–80%
Installation conditions
−
The floor at the installation location must be even, level and
have sufficient load-bearing capacity.
−
The device must be well illuminated.
−
The front of the device must be freely accessible.
Notes on hazardous substances
−
Ensure that the work room is sufficiently ventilated.
−
Adhere to emission limit values. If necessary, install an exhaust
air treatment system.
−
Do not direct drawn-off air back into the work room.
−
Install washing facilities and eye baths in the workplace.
−
Creeping vapors can constitute an ignition source. Seal the
room properly to ensure that gases or vapors which could
contain ignition sources cannot enter the room.
−
Install equipment for detecting and reporting gas hazards.
CAUTION
Environmental hazard
Hazardous substances, which could leak out as the
result of a fault during improperly performed
installation, can enter the ground.
For this reason:
– Install an impermeable base to protect against
hazardous substances used in the system.
Consult the safety data sheet of the hazardous
substance to determine which materials are
suitable for a drip tray or sealing.