Distillation Process Analyzer DPA-4.5
Commissioning
7.4 Monitoring and operation
Monitoring
•
After 2 hours of operation, check all connections for leaks and
make a general visual inspection (see page 129).
Supplying with product:
•
Check the setting on the system-side pressure regulator. If
necessary, re-adjust the pressure regulator.
•
Check the setting on the system-side flow meter. If necessary,
re-adjust the flow meter.
•
If necessary, repeat the adjustments to the pressure regulator
and flow meter until the target values are attained.
Operation
−
In normal operating circumstances, the analyzer works fully
automatically.
−
The analysis process can be followed on the analyzer’s
touchscreen.
−
The analyzer can be operated with the touchscreen stylus.
NOTE
For more detailed information on operating,
controlling and configuring the analyzer and on the
analyzing process, refer to the software handbook.
December 2013
107