Distillation Process Analyzer DPA-4.5
Technical data
3 Technical data
3.1 Measuring process
Specification
Device type
DPA-4.5
Variant „Low Pressure“ DPA-4.5
LowP
Measuring at reduced pressure
Measured variable
Distillation performance (distillate
quantity according to the temperature)
Measuring process
Batch distillation at atmospheric pressure
Measuring range
20 ... 420 °C (resolution 0.1 °C)
at 0 ... 100% condensed sample in the
receiver
Method
SAM (Standard
Analysis Mode)
RAM (Rapid
Analysis Mode)
ASTM D 86
DIN EN ISO 3405
IP 123
Similar to
Correlates with
Repeatability
Meets the requirements of the standards
Reproducibility
Meets the requirements of the standards
Cycle times (gasoline/diesel) information for T4
SAM (Standard
Analysis Mode)
RAM (Rapid
Analysis Mode)
IBP (min)
24/29
Depending on
product
Approx. 40% less
than SAM
50% point (min)
36/41
FBP (min)
45/50
December 2013