Maintenance and Troubleshooting 10-53
Problem
Individual PEAKVEL_1 or PEAKVEL_3 values
consistently read -9.99.
Possible
Causes
Sensor may be faulty.
5-volt fuse on regulator may have blown.
Possible
Solutions
Replace the Peak Combo Sensor or Surface Combo
Sensor.
Replace the 5-volt blue fuse (
when replaceable
) on the
regulator.
Caution:
Do not connect a questionable sensor to
another monitor before verifying the integrity of the
sensor. A shorted sensor will blow the 5-volt (blue)
replacement fuse on the regulator in the monitor.
Surface Velocity Subsystem
The following tables contain general troubleshooting techniques for
the surface velocity subsystem corresponding to the Surface Combo
Sensor.
Problem
Velocity readings are erratic or incorrect.
Possible
Causes
Sensor may be dirty or covered by debris.
Maximum Velocity
setting may require adjustment.
Transmit Frequency
setting may require adjustment.
Possible
Solutions
Clean the sensor transmit/receive surfaces.
Fine-tune Maximum Velocity setting in device
parameters.
Test other
Transmit Frequency
settings in the velocity
parameters.
Содержание Intrinsically-Safe TRITON+
Страница 19: ...1 12 ADS TRITON Manual Control Drawing ...
Страница 25: ...1 18 ADS TRITON Manual First page of the TRITON ATEX Certificate ...
Страница 27: ...1 20 ADS TRITON Manual First page of the TRITON IECEx Certificate of Conformity ...
Страница 29: ...1 22 ADS TRITON Manual CSA Certificate of Compliance Page 1 ...
Страница 30: ...Introduction 1 23 CSA Certificate of Compliance Page 2 ...
Страница 32: ...Introduction 1 25 Sample TRITON Declaration of Conformity Page 2 ...
Страница 34: ...Introduction 1 27 Activate Monitor Qstart software Run Diagnostics and Perform Confirmations Qstart software ...
Страница 49: ...2 12 ADS TRITON Manual ExPAC XBUS ...
Страница 159: ...External Power 5 9 Power supply firmly secured to the DIN rail Front view of power supply mounted to DIN rail ...
Страница 171: ...External Power 5 21 POWER SUPPLY DC OUTPUT TERMINAL CONNECTIONS Wires exiting DC output terminal on power supply ...
Страница 231: ...6 44 ADS TRITON Manual Editing the Smart Ultrasonic Depth Sensor Device Edit Smart Depth 1 Parameters dialog ...