Maintenance and Troubleshooting 10-19
Replace the dryer tube or the desiccant contained within the dryer
tube on a regular basis or as necessary. Desiccant can absorb
moisture only when it is blue in color. Pink desiccant indicates that
it can no longer absorb moisture. Therefore, if the desiccant
appears either colorless or pink upon inspection, replace the dryer
tube or desiccant immediately.
Replacing the Dryer Tube
Replace the pressure depth sensor dryer tube (ADS p/n 3704-0032)
in the following way:
1.
Clip the cable ties securing the dryer tube to the monitor.
2.
Cut the clear flexible tubing running from the dryer tube to the
pressure depth sensor connector at a location close to the brass
barbed fitting on the dryer tube.
3.
Set aside, but do not discard, the old dryer tube with the used
desiccant.
4.
Attach a new dryer tube to the pressure depth sensor by
inserting the brass barbed fitting into the open end of the plastic
tubing running from the sensor connector. Make sure the
tubing seats firmly against the fitting to prevent air or moisture
transfer.
Inserting the brass barbed fitting into the tube
5.
Secure the new dryer tube to the monitor.
Содержание Intrinsically-Safe TRITON+
Страница 19: ...1 12 ADS TRITON Manual Control Drawing ...
Страница 25: ...1 18 ADS TRITON Manual First page of the TRITON ATEX Certificate ...
Страница 27: ...1 20 ADS TRITON Manual First page of the TRITON IECEx Certificate of Conformity ...
Страница 29: ...1 22 ADS TRITON Manual CSA Certificate of Compliance Page 1 ...
Страница 30: ...Introduction 1 23 CSA Certificate of Compliance Page 2 ...
Страница 32: ...Introduction 1 25 Sample TRITON Declaration of Conformity Page 2 ...
Страница 34: ...Introduction 1 27 Activate Monitor Qstart software Run Diagnostics and Perform Confirmations Qstart software ...
Страница 49: ...2 12 ADS TRITON Manual ExPAC XBUS ...
Страница 159: ...External Power 5 9 Power supply firmly secured to the DIN rail Front view of power supply mounted to DIN rail ...
Страница 171: ...External Power 5 21 POWER SUPPLY DC OUTPUT TERMINAL CONNECTIONS Wires exiting DC output terminal on power supply ...
Страница 231: ...6 44 ADS TRITON Manual Editing the Smart Ultrasonic Depth Sensor Device Edit Smart Depth 1 Parameters dialog ...