D-1
A P P E N D I X D
System Configuration and Setup to
Support the Telog Ru-33
Using a battery-powered ADS
®
™
to support the Telog
®
Ru-33 Recording Telemetry Unit (RTU) requires some limited
configuration and setup. These procedures involve designating the
appropriate identification information to enable recognition and
data exchange between the ADS and Telog units and using an
interface cable to connect the monitor to the RTU.
Note:
The Telog Ru-33 must have firmware version
3.300 or later to effectively request data from the
monitor.
Connecting the units together primarily involves running the ADS
Triton-Telog Comm Cable from the battery-powered Ru-33 to the
battery-powered
monitor. This application requires
installing both units in the manhole, a hazardous environment.
However, ADS offers an optional extension cable (ADS p/n 8000-
0044-25) that allows you to install the Telog unit outside the
hazardous environment.
Note:
The Telog Ru-33 does not possess IS (intrinsically-
safe) certification. Therefore, an installation involving
connecting the Telog to the IS
with both units
residing
inside the manhole
will
not
be considered
intrinsically safe. The installation will be considered
Содержание Intrinsically-Safe TRITON+
Страница 19: ...1 12 ADS TRITON Manual Control Drawing ...
Страница 25: ...1 18 ADS TRITON Manual First page of the TRITON ATEX Certificate ...
Страница 27: ...1 20 ADS TRITON Manual First page of the TRITON IECEx Certificate of Conformity ...
Страница 29: ...1 22 ADS TRITON Manual CSA Certificate of Compliance Page 1 ...
Страница 30: ...Introduction 1 23 CSA Certificate of Compliance Page 2 ...
Страница 32: ...Introduction 1 25 Sample TRITON Declaration of Conformity Page 2 ...
Страница 34: ...Introduction 1 27 Activate Monitor Qstart software Run Diagnostics and Perform Confirmations Qstart software ...
Страница 49: ...2 12 ADS TRITON Manual ExPAC XBUS ...
Страница 159: ...External Power 5 9 Power supply firmly secured to the DIN rail Front view of power supply mounted to DIN rail ...
Страница 171: ...External Power 5 21 POWER SUPPLY DC OUTPUT TERMINAL CONNECTIONS Wires exiting DC output terminal on power supply ...
Страница 231: ...6 44 ADS TRITON Manual Editing the Smart Ultrasonic Depth Sensor Device Edit Smart Depth 1 Parameters dialog ...