6-80 ADS Manual
5.
Enter the distance measured manually from the bottom of the
pipe to the flow surface in the
Depth
field
or
the range from
the pipe crown/sensor face (as applicable) to the flow surface in
the
Range
field. The measurement perspective depends on the
option selected in step 4. Enter any estimated deviation
corresponding to the measurement in the
+/-
field.
6.
Enter the velocity of the flow measured manually using a
portable velocity meter in the
Velocity
field.
7.
Enter the initials or designation for the field personnel in the
Crew
field.
8.
Enter any information corresponding to the field measurements,
the flow conditions, or location that may be relevant to the
confirmations in the
Comments
field.
9.
Click on the
Read
button to prompt the monitor to take depth
and velocity measurements.
The Monitor Measurement section displays the depth and
velocity readings taken by the monitor.
10.
Compare the manual field measurements to the sensor readings.
If a difference exists between the manual depth measurements
and the depth sensor measurements, you can compensate for
this difference by editing the
electronic offset
parameter for the
appropriate depth sensor. Consider applying this offset to
ensure the monitor produces the most accurate data. Use the
following equation to determine the electronic offset:
Field Depth = Sensor Depth + Electronic Offset
11.
If the sensor reading is
low
, add a
positive
(+) electronic offset.
12.
If the sensor reading is
high
, add a
negative
(-) electronic
offset.
Note:
Exercise caution when applying an electronic
offset. Use an offset only when you are confident in the
accuracy and reliability of the manual measurements and
physical offset of the sensor.
Содержание Intrinsically-Safe TRITON+
Страница 19: ...1 12 ADS TRITON Manual Control Drawing ...
Страница 25: ...1 18 ADS TRITON Manual First page of the TRITON ATEX Certificate ...
Страница 27: ...1 20 ADS TRITON Manual First page of the TRITON IECEx Certificate of Conformity ...
Страница 29: ...1 22 ADS TRITON Manual CSA Certificate of Compliance Page 1 ...
Страница 30: ...Introduction 1 23 CSA Certificate of Compliance Page 2 ...
Страница 32: ...Introduction 1 25 Sample TRITON Declaration of Conformity Page 2 ...
Страница 34: ...Introduction 1 27 Activate Monitor Qstart software Run Diagnostics and Perform Confirmations Qstart software ...
Страница 49: ...2 12 ADS TRITON Manual ExPAC XBUS ...
Страница 159: ...External Power 5 9 Power supply firmly secured to the DIN rail Front view of power supply mounted to DIN rail ...
Страница 171: ...External Power 5 21 POWER SUPPLY DC OUTPUT TERMINAL CONNECTIONS Wires exiting DC output terminal on power supply ...
Страница 231: ...6 44 ADS TRITON Manual Editing the Smart Ultrasonic Depth Sensor Device Edit Smart Depth 1 Parameters dialog ...