10-4 ADS Manual
Make sure all connector ports on the monitor top are dry. Use
compressed air or paper towels to dry out wet or damp
connectors. Note and replace missing interfacial seals in the
sensor connector ends. As a temporary measure, the
connectors can be protected against moisture by wrapping with
self-vulcanizing rubber stretch tape. Make sure covers are
securely in place on any unused ports on the monitor top.
Confirming the Monitor
Confirm the accuracy of the sensor subsystems in the monitor on a
regular basis through
Qstart
. Confirmation involves comparing
manually-measured depth of flow and velocity readings to the
monitor's readings. This process also verifies sensor parameters.
Refer to the
Qstart
online help for detailed information on
performing confirmation procedures.
Obtaining Diagnostic Codes from the Monitor
The
enables you to observe monitor activity, view
diagnostic information, and confirm test cryouts from the monitor
through a small LED status window on the monitor top. The
numbers/characters displayed represent specific codes indicating
current processes, identification, status, or readings associated with
the system and communications. The following sections contain
general descriptions and interpretations of the diagnostic
information available through the LED and instructions for
accessing information or prompting a message.
Содержание Intrinsically-Safe TRITON+
Страница 19: ...1 12 ADS TRITON Manual Control Drawing ...
Страница 25: ...1 18 ADS TRITON Manual First page of the TRITON ATEX Certificate ...
Страница 27: ...1 20 ADS TRITON Manual First page of the TRITON IECEx Certificate of Conformity ...
Страница 29: ...1 22 ADS TRITON Manual CSA Certificate of Compliance Page 1 ...
Страница 30: ...Introduction 1 23 CSA Certificate of Compliance Page 2 ...
Страница 32: ...Introduction 1 25 Sample TRITON Declaration of Conformity Page 2 ...
Страница 34: ...Introduction 1 27 Activate Monitor Qstart software Run Diagnostics and Perform Confirmations Qstart software ...
Страница 49: ...2 12 ADS TRITON Manual ExPAC XBUS ...
Страница 159: ...External Power 5 9 Power supply firmly secured to the DIN rail Front view of power supply mounted to DIN rail ...
Страница 171: ...External Power 5 21 POWER SUPPLY DC OUTPUT TERMINAL CONNECTIONS Wires exiting DC output terminal on power supply ...
Страница 231: ...6 44 ADS TRITON Manual Editing the Smart Ultrasonic Depth Sensor Device Edit Smart Depth 1 Parameters dialog ...