3-14 ADS
Manual
5.
Slide the Surface Combo/Ultrasonic Depth Sensor mounting
plate onto the open end of the ring with the back of the
mounting plate (side with the slots) facing the outside of the
ring. The side with the keyholes should face the inside of the
ring. The end of the plate with the keyholes should be
extending out from the
upstream
edge of the ring (
opposite
the
edge with the holes for securing the sensor cable).
Sliding the Surface Combo/Ultrasonic Depth Sensor mounting plate onto the
ring
Note:
For applications that involve replacing a
previous
version of the Ultrasonic Depth Sensor (for example, ADS
p/n 8K-CS3-V0-XX-30) with a
new
Ultrasonic Depth
Sensor (ADS p/n 8K-CS5-D1-00-30) or Surface Combo
Sensor and an existing ring remains assembled and secured
in the pipe, use a sliding adapter bracket (ADS p/n 8000-
0299) to accommodate the new sensor in the existing
sensor mounting plate. Once the previous sensor has been
Содержание Intrinsically-Safe TRITON+
Страница 19: ...1 12 ADS TRITON Manual Control Drawing ...
Страница 25: ...1 18 ADS TRITON Manual First page of the TRITON ATEX Certificate ...
Страница 27: ...1 20 ADS TRITON Manual First page of the TRITON IECEx Certificate of Conformity ...
Страница 29: ...1 22 ADS TRITON Manual CSA Certificate of Compliance Page 1 ...
Страница 30: ...Introduction 1 23 CSA Certificate of Compliance Page 2 ...
Страница 32: ...Introduction 1 25 Sample TRITON Declaration of Conformity Page 2 ...
Страница 34: ...Introduction 1 27 Activate Monitor Qstart software Run Diagnostics and Perform Confirmations Qstart software ...
Страница 49: ...2 12 ADS TRITON Manual ExPAC XBUS ...
Страница 159: ...External Power 5 9 Power supply firmly secured to the DIN rail Front view of power supply mounted to DIN rail ...
Страница 171: ...External Power 5 21 POWER SUPPLY DC OUTPUT TERMINAL CONNECTIONS Wires exiting DC output terminal on power supply ...
Страница 231: ...6 44 ADS TRITON Manual Editing the Smart Ultrasonic Depth Sensor Device Edit Smart Depth 1 Parameters dialog ...