Maintenance and Troubleshooting 10-47
Upward Depth Subsystem
The following tables contain general troubleshooting techniques for
the upward depth subsystem corresponding to the Peak Combo
Sensor.
Note:
Contact your regional ADS representative for
further diagnosis prior to replacing a sensor.
Problem
UpDepth data periodically reads 0.63 inches (16 mm).
Possible
Causes
Minimum flows are occurring within sensor deadband
(extending 1.0 inches (25 mm) up from bottom of Peak
Combo Sensor) or the sensor is receiving a strong
return signal close to the sensor face.
Sensor transmit/receive surfaces are dirty.
Sensor or cabling might be faulty.
Possible
Solutions
Verify minimum flow depths are not occurring within
sensor deadband.
Set the deadband on the UpDepth sensor to a setting
other than
Auto
, but well below the expected minimum
flow depth. Fire the sensor several times to verify
whether the readings improve.
Ensure the sensor is installed as closely as possible to
the 6:00 position in the pipe, silt allowing.
Clean the sensor transmit/receive surfaces.
Inspect the sensor cables for integrity.
Check for moisture in the connector.
Contact your regional ADS representative.
Replace the sensor.
Replace the monitor if defective.
Содержание Intrinsically-Safe TRITON+
Страница 19: ...1 12 ADS TRITON Manual Control Drawing ...
Страница 25: ...1 18 ADS TRITON Manual First page of the TRITON ATEX Certificate ...
Страница 27: ...1 20 ADS TRITON Manual First page of the TRITON IECEx Certificate of Conformity ...
Страница 29: ...1 22 ADS TRITON Manual CSA Certificate of Compliance Page 1 ...
Страница 30: ...Introduction 1 23 CSA Certificate of Compliance Page 2 ...
Страница 32: ...Introduction 1 25 Sample TRITON Declaration of Conformity Page 2 ...
Страница 34: ...Introduction 1 27 Activate Monitor Qstart software Run Diagnostics and Perform Confirmations Qstart software ...
Страница 49: ...2 12 ADS TRITON Manual ExPAC XBUS ...
Страница 159: ...External Power 5 9 Power supply firmly secured to the DIN rail Front view of power supply mounted to DIN rail ...
Страница 171: ...External Power 5 21 POWER SUPPLY DC OUTPUT TERMINAL CONNECTIONS Wires exiting DC output terminal on power supply ...
Страница 231: ...6 44 ADS TRITON Manual Editing the Smart Ultrasonic Depth Sensor Device Edit Smart Depth 1 Parameters dialog ...