3-56 ADS
Manual
Securing the other end of the band
Note:
If the mounting plate has too much side play, attach
another anchor bolt approximately 1 inch (22 mm) from
the plate.
10.
Slide the sensor mounting plate along the metal band until it is
level (from side to side) with the flow surface and pipe crown,
and tighten the nuts securely with a nut driver. For more
information, refer to the instructions on leveling this sensor in
Installing the Ring in the Pipe
beginning on page 3-29.
11.
Install the Surface Combo Sensor/Ultrasonic Depth Sensor onto
the mounting plate by inserting the studs on the back of the
sensor into the keyholes on the mounting plate and sliding the
sensor back (in a
downstream
direction) until it is fully in
place. The sensor cable should exit the downstream edge of the
band.
Содержание Intrinsically-Safe TRITON+
Страница 19: ...1 12 ADS TRITON Manual Control Drawing ...
Страница 25: ...1 18 ADS TRITON Manual First page of the TRITON ATEX Certificate ...
Страница 27: ...1 20 ADS TRITON Manual First page of the TRITON IECEx Certificate of Conformity ...
Страница 29: ...1 22 ADS TRITON Manual CSA Certificate of Compliance Page 1 ...
Страница 30: ...Introduction 1 23 CSA Certificate of Compliance Page 2 ...
Страница 32: ...Introduction 1 25 Sample TRITON Declaration of Conformity Page 2 ...
Страница 34: ...Introduction 1 27 Activate Monitor Qstart software Run Diagnostics and Perform Confirmations Qstart software ...
Страница 49: ...2 12 ADS TRITON Manual ExPAC XBUS ...
Страница 159: ...External Power 5 9 Power supply firmly secured to the DIN rail Front view of power supply mounted to DIN rail ...
Страница 171: ...External Power 5 21 POWER SUPPLY DC OUTPUT TERMINAL CONNECTIONS Wires exiting DC output terminal on power supply ...
Страница 231: ...6 44 ADS TRITON Manual Editing the Smart Ultrasonic Depth Sensor Device Edit Smart Depth 1 Parameters dialog ...